The Air-Shields TI-100 Transport Incubator is available New from Med Vest Technology.

FEATURES

  • Intended for transport of high risk, premature, low birth-weight, or critically ill newborns.
  • It provides a means to control air temperature, oxygen concentrations, and relative humidity.
  • A double wall hood permits full visibility and provides an effective thermal barrier from the environment.

 

SPECIFICATIONS

Dimensions (with stand):

  • Length: 40.25 in. (102 cm)
  • Width: 22.25 in. (56.5 cm)
  • Height: 46.75 in. (118.7 cm) maximum, 34.75 in. (88.3 cm) minimum

Power Requirements:

  • AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
  • DC: 12 VDC, 115 Watts

Internal Battery:

  • Type: gel type, sealed and rechargeable.
  • Voltage: 12 VDC nominal
  • Quantity: 2
  • Charging Time: two batteries take 20 minutes to recharge from fully discharged.
  • Life expectancy: minimum 200 full charge/discharge cycles.

Chassis Leakage Current: 100 µA or less

Alarms:

  • High Temp: Actuates at displayed incubator temperature of 39° ± 0.5 °C.
  • Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
  • Power Failure: Actuates if main AC power fails and no DC back-up power is present.
  • Heater Temp: Actuates if heater temperature exceeds 77 °C nominal.
  • Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.

Temperature:

  • Set Point Range: 21.5 °C ± 1.0 °C to 38 °C
  • Rise Time: 30 minutes
  • Variability: 0.2 °C
  • Overshoot: 1.5 °C maximum
  • Uniformity: 1 °C
  • Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1 °C
  • Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5 °C
  • Observation Light Output: 10-foot candles at mattress level

 

-->