Critikon 8720 Multi-Parameter Monitor

Critikon 8720 Multi-Parameter Monitor

Web ID: 955
Condition: Brand New

0 item is in your cart. View cart now

Need help choosing?

Ask an Equipment Expert

(781) 302-6019

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Features

Now the DINAMAP® Compact Monitor is available in four different configurations offering you the technology that's right for you.

Each model features our unique DINAMAP® Technology, the "Gold Standard" in BP accuracy. Optional, integrated printers are also available in each model.

  • Suitable for Adult, Pediatric and Neonatal Use
  • Maintains Accuracy even when the Patient Moves
  • Allows you to take all Connected Vitals with the Press of One Button
  • STAT Mode for Rapid BP Readings in Emergency Situations
  • Single Determinations or Automatic Monitoring of Blood Pressure
  • User-Adjustable Target Inflation Pressure
  • Oral / Rectal Predictive Temperature
  • Built-In Printer Available

Specifications

Noninvasive Blood Pressure

  • Heart Rate Range: 10-250 bpm
  • Connector: AAMI recommended, ECGC; MS3102A-14S-6S
  • Leads: Selectable, I, II, III Lead off detection
  • Band Width: 0.5-40 Hz
  • Sweep Speed: 12.5, 25 and 50 mm per second
  • Gain: 1mV to 8mV full scale, selectable in 5 steps
  • QRS Detection Tone: Volume selectable from 0(off) to 15
  • Alarm Limits: Heart rate-high, 80-250 bpm; low, 10-140 bpm

ECG

  • Heart Rate Range: 10-250 bpm
  • Connector: AAMI recommended, ECGC; MS3102A-14S-6S
  • Leads: Selectable, I, II, III Lead off detection
  • Band Width: 0.5-40 Hz
  • Sweep Speed: 12.5, 25 and 50 mm per second
  • Gain: 1mV to 8mV full scale, selectable in 5 steps
  • QRS Detection Tone: Volume selectable from 0(off) to 15
  • Alarm Limits: Heart rate-high, 80-250 bpm; low, 10-140 bpm

Oxygen Saturation

  • Accuracy: (± standard deviation) ±2% at 90-99%; ±3% at 75-89%; unspecified under 75%
  • Saturation Range: 40-99%
  • Pulse Rate: From 30-250 bpm
  • Pulse Averaging: User selectable for 0 to 30 sec. in increments of 3. Preset at 6 seconds at power turn on.
  • Saturation Pitch Indicator: Pitch changes with saturation. Volume selectable from 0(off) to 15.
  • Alarm Limits: Saturation - high, 90-99%; low, 40-95%

Alarm Functions

  • System: Internal system test performed at power-up and during operation
  • Volume: Adjustable from 1 to 15 on all parameters
  • NIBP: Procedural alarm for overpressure, excess determination time, excess pump up time, no determination
  • SpO2: Procedural alarms for low signal, probe disconnect, probe failure

Mechanical

  • Weight: 9.8 lbs
  • Dimensions: 9.25 (H) × 720 (W) × 8.5 (D) inches
  • Display: Electroluminescent, 320 × 256 pixels; viewing area, 3.0 (H) × 3.8 (W) inches
  • Power Requirements: 88-268 VAC; 47-63 Hz
  • Fuse Requirements: 2.0 amps
  • Leakage Current: <100µA
-->