Colin BP-S510 Patient Monitor

Colin BP-S510 Patient Monitor

Brand: Colin
Type: Monitors
Web ID: 1228
Condition: Brand New
Product Number: 65510

0 item is in your cart. View cart now

Need help?

Get an Equipment Quote

(781) 302-6019

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Leading-edge patient monitor features patented Colin NIBP technology

The BP-S510 has a sophisticated design that provides simple, user-friendly operation. Its advanced functionality makes it perfect for a wide variety of uses, including post-anesthesia care, ambulatory surgery, endoscopy, emergency rooms and more!

  • Measures — ECG (3/5 lead), heart rate, Colin NIBP, Nellcor SpO2 and respiration. Available with temperature, two-channel IBP, EtCO2, and a printer.
  • Compact — Conveniently compact and feature-packed, the BP-S510 is great for areas where space is at a premium.
  • Lightweight — Weighing only 12.1 lbs, the BP-S510 is easy to carry from ward to ward. This lightweight device makes it easy to move to stepdown, PACU, RR, etc.
  • Battery — Operation for Portability: Available with rechargeable batteries, the BP-S510 can be used during transport within your facility.
  • Wide Viewing Angle — The BP-S510's large LCD screen features a wide viewing angle. Each parameter is color-coded in bright and vivid colors, making the information easy to read even from a distance. You can also select from multiple views to highlight just the information you need.

Specifications

Non-Invasive Blood Pressure (NIBP) - Patented Colin Technology

  • Measurement technology Oscillometric method
  • Measurement method Linear deflation method
  • NIBP Measurement range
  • Adult/pediatric mode:
    • SYS 60 to 250mmHg
    • MAP 45 to 235mmHg
    • DIA 40 to 200mmHg
    • Pulse rate 40 to 200bpm
  • Pressure display range 10 to 300mmHg
  • Pressure display accuracy Less than ±3mmHg
  • Pulse rate 40 to 240bpm
  • NIBP accuracy AAMI SP-10:2002+A1:2003
  • Pulse rate accuracy ±2% or ±2 beats

ECG

  • Leads 3/5 Lead
  • Lead I, II, III, aVR, aVL, aVF, V (Chest Lead)
  • Heart Rate
  • Measurement Range 20 to 250bpm
  • Accuracy ±3bpm or ±5% (whichever is greater)

Respiration (im RR)

  • Measurement method Trans-thoracic impedance
  • Leads RA to LA
  • Range 0, 3 to 120 breaths/min
  • Accuracy ±3 breaths/min

SpO2 (Pulse oximeter)

  • Measurement method 2 wave length pulse wave type
  • Range 1 to 100% SpO2
  • Accuracy Adult/pediatric (When using SpO2 Disposable Sensor MAX-A) 70 to 100% SpO2 ±2 digits, 1 to 69% SpO2 unspecified
  • Pulse Rate Range 20 to 250bpm
  • Pulse Rate Accuracy ±3bpm
  • Display update Less than 30 sec

Temperature Measurement (Optional)

  • Measurement method Thermistor
  • Probe Type Thermistor probe YSI 400 series (2 channels) (Probes not included)
  • Range 15 to 45°C (59 to 113°F)
  • Display Accuracy ±0.1°C (25 to 45°C) (77 to 113°F) ±0.2°C (15 to less than 25°C) (59 to less than 77°F)
  • Probe Accuracy ±0.1°C (±0.18°F)

IBP (Optional)

  • Measurement Method: Pressure Transducer Method
  • Pressure Measurement Range - -50 to +300mmHG
  • Pulse Rate Measurement Range – 20 to 250bpm
  • Display Accuracy - ±3mmHG
  • Standard – IEC60601-2-34:2000

CO2 (Option)

  • Measurement method Non-dispersive Infrared Spectroscopy
  • Range 0-100mmHg : 0-13.3kpa : 0-10%CO2
  • Accuracy ±2mmHg or ±4% (whichever is greater)
  • Respiration (awRR) Range 0 to 120 breaths/min
  • Respiration (awRR) Accuracy ±1 breath/min

Recorder (Option)

  • Type Thermal
  • Weight 0.33 lbs. (150 g)
  • Resolution 8 dot/mm
  • Number of channels 1 to 2 channels
  • Paper Type Thermal
  • Paper Width 2 inches (50 mm)
  • Paper Speeds 25.0mm/sec and 50.0mm/sec (1in/sec and 2in/sec)

Power Requirement

  • AC Power 100-240V~50/60Hz, 0.5-1.3A
  • Battery Ni-MH, 12V/3.8Ah (1 included)

Environmental

  • Operating Temperature 10 to 40°C (50 to 104°F)
  • Storage Temperature -20 to 60°C (-4 to 140°F)
  • Operating Humidity 15 to 90% RH, non-condensing
-->